Abstract

This study evaluates the inter-device measurement properties of partial coherence interferometry (PCI) and spectral domain optical coherence tomography (SD-OCT) in measuring axial length, particularly for myopia management. We recruited 82 eyes from 41 adult participants with a mean age of 31.0 ± 17.6 years and a mean spherical equivalent of −2.20 ± 2.28 D. Axial length was measured using SD-OCT and PCI for both the right and left eyes. Agreement between the two measurements was assessed using Bland–Altman analysis, and graphs and values were compared with linear mixed models. The results show a near-to-zero and non-significant bias between measurements. The 95% limits of agreement showed a value of 0.06 mm. Both devices can accurately measure the axial length. OCT biometry performed with SD-OCT can be successfully interchanged with partial coherence interferometry, but they should be cautiously interchanged when performing longitudinal comparisons.

Original languageEnglish
Article number46
Pages (from-to)1-8
JournalVision (Switzerland)
Volume8
Issue number3
DOIs
Publication statusPublished - Sept 2024

OECD Field of Science

  • 1.3 Physical Sciences

Keywords

  • biometry
  • OCT
  • myopia
  • partial coherence interferometry
  • axial length

Fingerprint