Evaluating the quality of implantation of percutaneous ventricular restoration device (Parachute®) by cardiac computed tomography

  • Mohamad Amer Alaiti
  • , Anas Fares
  • , Andrejs Erglis
  • , Lemba Nshisso
  • , Kashif Shaikh
  • , Anthony E. DeCicco
  • , Marwan Nasif
  • , Ahmad Alkhalil
  • , Hüseyin Ince
  • , William T. Abraham
  • , Daniel I. Simon
  • , Marco A. Costa
  • , Guilherme F. Attizzani
  • , Hiram G. Bezerra*
  • *Corresponding author for this work
  • Case Western Reserve University
  • Paula Stradina Clinical University Hospital
  • University of Rostock
  • Ohio State University

Research output: Contribution to journalArticlepeer-review

2 Citations (Scopus)

Abstract

Background: The Parachute is a novel percutaneously implanted ventricular partitioning device (VPD) that has emerged as a safe and feasible treatment option for patients with heart failure following anterior wall myocardial infarction. VPD efficacy is likely dependent on optimal device placement, but to date there are no published data examining the effect of device positioning on patient outcomes. Methods and results: We retrospectively identified 32 patients successfully implanted with the Parachute device, all of whom underwent cardiac computed tomography (CCT) at baseline and after 6 months of follow-up. Patients were divided into two groups based on self-reported improvement in New York Heart Association (NYHA) functional class: “not improved NYHA” (n = 12) and “improved NYHA” (n = 20). There were significant differences between both groups with regard to device positioning on follow-up CCT. Compared to patients with “improved NYHA,” patients with “not improved NYHA” had longer distances from device foot to left ventricular apex (8.0 ± 4.9 vs. 2.9 ± 4.6 mm; P = 0.01), and higher lateral angles (18.0 ± 14 vs. 9.1 ± 6.8 degrees; P = 0.02), respectively. There was no significant difference between the two groups in landing zone (45.4 ± 7. vs. 45.1 ± 6.9 mm; P = 0.92) and inferior angle (14.0 ± 11.9 vs. 14.3 ± 10.1 degrees; P = 0.95). There was a numerically larger malapposition area in the “not improved NYHA” group (5.1 ± 4.5 vs. 3.2 ± 2.2 cm2; P = 0.12). Conclusion: Quality of Parachute implant impacted clinical outcome, these findings should be applied prospectively in helping operators to achieve optimal implant.

Original languageEnglish
Pages (from-to)E104-E111
JournalCatheterization and Cardiovascular Interventions
Volume89
Issue number4
DOIs
Publication statusPublished - 1 Mar 2017
Externally publishedYes

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Keywords

  • electron beam CT/multidetector CT
  • heart failure
  • imaging
  • interventional devices/innovation

Fingerprint

Dive into the research topics of 'Evaluating the quality of implantation of percutaneous ventricular restoration device (Parachute®) by cardiac computed tomography'. Together they form a unique fingerprint.

Cite this