Abstract
We present preliminary results of X-ray measurements on two small format compound semiconductor arrays. The devices, a 5×5 gallium arsenide array and a 3×3 cadmium zinc telluride array, were produced specifically to address the material, electronic and technological problems that need to be solved in order to develop mega-pixel, Fano limited spectroscopic arrays. The GaAs device was fabricated on 40 μm epitaxial material and has a pixel size of 200×200 microns2 with pitch 250 μm. The CdZnTe array was fabricated on a 5×5×1.6 mm3 single crystal of spectroscopic quality. The pixel sizes were 350×350 microns2 with a pixel pitch of 250 μm. Measurements from 5.9 keV to 100 keV were carried out both in our laboratory and at the HASYLAB synchrotron research facility in Hamburg, Germany. The typical FWHM energy resolutions recorded at 5.9 keV by the GaAs and CdZnTe arrays were 394 eV and 900 eV, respectively.
| Original language | English |
|---|---|
| Pages (from-to) | 55-66 |
| Number of pages | 12 |
| Journal | Proceedings of SPIE - The International Society for Optical Engineering |
| Volume | 4141 |
| Issue number | 1 |
| DOIs | |
| Publication status | Published - 21 Nov 2000 |
| Externally published | Yes |