@inproceedings{1cfca78d15434827b6f2fd27acf84734,
title = "How Close Can You Go? Minimum Eye-to-Screen Distance for EyeLink 1000 Plus",
abstract = "Certain research paradigms require close eye-to-screen distances. EyeLink 1000 Plus is one of the most widely used eye trackers, yet there is no documented guidance on the minimum eye-to-screen distance that still allows successful calibration. We tested three lenses (16mm, 25mm, 35mm) to determine this minimum while maintaining at least 15° of visual angle in the smallest dimension. The 16mm and 25mm lenses allowed the closest distances (330{\textendash}335 mm to screen top), while the 35mm lens required 457 mm. All configurations achieved average validation accuracy under 0.30°.",
keywords = "Eye movements, Closest distance, Eye tracking, Visual angle",
author = "Mohammadhossein Salari and Renārs Truk{\v s}a and Roman Bednarik",
note = "Publisher Copyright: {\textcopyright} 2026 Copyright held by the owner/author(s).; 18th ACM Symposium on Eye Tracking Research and Applications, ETRA 2026 ; Conference date: 01-06-2026 Through 04-06-2026",
year = "2026",
month = may,
day = "31",
doi = "10.1145/3797246.3805851",
language = "English",
series = "ETRA 2026 - Proceedings of the 2026 Symposium on Eye Tracking Research and Applications",
publisher = "Association for Computing Machinery, Inc",
pages = "1--3",
editor = "Aladine Chetouani and Arantxa Vilanueva and Alessandro Bruno and Rebeca Echeverria",
booktitle = "Proceedings of the 2026 Symposium on Eye Tracking Research and Applications (ETRA 2026)",
}