Optical coherence tomography versus intravascular ultrasound to evaluate coronary artery disease and percutaneous coronary intervention

  • Hiram G. Bezerra
  • , Guilherme F. Attizzani
  • , Vasile Sirbu
  • , Giuseppe Musumeci
  • , Nikoloz Lortkipanidze
  • , Yusuke Fujino
  • , Wei Wang
  • , Sunao Nakamura
  • , Andrej Erglis
  • , Giulio Guagliumi
  • , Marco A. Costa*
  • *Šī darba korespondējošais autors
  • Case Western Reserve University
  • Ospedali Riuniti di Bergamo
  • New Tokyo Hospital
  • Paula Stradina Clinical University Hospital

Zinātniskās darbības rezultāts: Devums žurnālamZinātniskais raksts (žurnālā)koleģiāli recenzēts

153 Atsauces (Scopus)

Kopsavilkums

Objectives We compared intravascular ultrasound (IVUS) and 2 different generations of optical coherence tomography (OCT)-time-domain OCT (TD-OCT) and frequency-domain OCT (FD-OCT)-for the assessment of coronary disease and percutaneous coronary intervention (PCI) using stents. Background OCT is a promising light-based intravascular imaging modality with higher resolution than IVUS. However, the paucity of data on OCT image quantification has limited its application in clinical practice. Methods A total of 227 matched OCT and IVUS pull backs were studied. One hundred FD-OCT and IVUS pull backs in nonstented (n = 56) and stented (n = 44) vessels were compared. Additionally, 127 matched TD-OCT and IVUS images were compared in stented vessels. Results FD-OCT depicted more severe native coronary disease than IVUS; minimal lumen area (MLA) was 2.33 ± 1.56 mm2 versus 3.32 ± 1.92 mm2, respectively (p < 0.001). Reference vessel dimensions were equivalent between FD-OCT and IVUS in both native and stented coronaries, but TD-OCT detected smaller reference lumen size compared with IVUS. Immediately post-PCI, in-stent MLAs were similar between FD-OCT and IVUS, but at follow-up, both FD-OCT and TD-OCT detected smaller MLAs than did IVUS, likely due to better detection of neointimal hyperplasia (NIH). Post-PCI malapposition and tissue prolapse were more frequently identified by FD-OCT. Conclusions FD-OCT generates similar reference lumen dimensions but higher degrees of disease severity and NIH, as well as better detection of malapposition and tissue prolapse compared with IVUS. First-generation TD-OCT was associated with smaller reference vessel dimensions compared with IVUS.

OriģinālvalodaAngļu
Lapas (no-līdz)228-236
Lapu skaits9
ŽurnālsJACC: Cardiovascular Interventions
Sējums6
Izdevuma numurs3
DOIs
Publikācijas statussPublicēts - marts 2013

ANO IAM

Šis izpildes rezultāts palīdz sasniegt šādus ANO ilgtspējīgas attīstības mērķus (IAM)

  1. 3. IAM — Laba Veselība un Labbūtība
    3. IAM — Laba Veselība un Labbūtība

OECD Zinātnes nozare

  • 3. Medicīnas un veselības zinātnes

Nospiedums

Uzziniet vairāk par pētniecības tēmām “Optical coherence tomography versus intravascular ultrasound to evaluate coronary artery disease and percutaneous coronary intervention”. Kopā tie veido unikālu nospiedumu.

Citēt šo