Primary endpoint results of the OMEGA Study: One-year clinical outcomes after implantation of a novel platinum chromium bare metal stent

  • John C. Wang*
  • , Didier Carrié
  • , Monica Masotti
  • , Andrejs Erglis
  • , David Mego
  • , Matthew W. Watkins
  • , Paul Underwood
  • , Dominic J. Allocco
  • , Christian W. Hamm
  • *Šī darba korespondējošais autors
  • MedStar Health
  • CHU de Toulouse
  • University of Barcelona
  • Paula Stradina Clinical University Hospital
  • Arkansas Heart Hospital
  • University of Vermont
  • Boston Scientific Corporation
  • Kerckhoff Heart and Thorax Center

Zinātniskās darbības rezultāts: Devums žurnālamZinātniskais raksts (žurnālā)koleģiāli recenzēts

15 Atsauces (Scopus)

Kopsavilkums

Background/purpose: Bare metal stents (BMS) have similar rates of death and myocardial infarction (MI) compared to drug-eluting stents (DES). DES lower repeat revascularization rates compared to BMS, but may have higher rates of late stent thrombosis (ST) potentially due to impaired endothelialization requiring longer dual anti-platelet therapy (DAPT). OMEGA evaluated a novel BMS designed to have improved deliverability and radiopacity, in comparison to currently available platforms. Methods/materials: OMEGA was a prospective, multicenter, single-arm study enrolling 328 patients at 37 sites (US and Europe). Patients received the OMEGA stent (bare platinum chromium element stent) for the treatment of de novo native coronary artery lesions (≤28 mm long; diameter ≥2.25 mm to ≤4.50 mm). The primary endpoint was 9-month target lesion failure (TLF: cardiac death, target vessel-related MI, target lesion revascularization [TLR]) compared to a prespecified performance goal (PG) based on prior generation BMS. All major cardiac events were independently adjudicated. DAPT was required for a minimum of 1. month post procedure. Results: In the OMEGA study, the mean age was 65; 17% had diabetes mellitus. The primary endpoint was met; 9. month TLF rate was 11.5%, and the upper 1-sided 95% confidence bound of 14.79% was less than the prespecified PG of 21.2% (p < 0.0001). One-year event rates were low including a TLF rate of 12.8% and an ST rate of 0.6% at 12. months. Conclusions: One-year outcomes of OMEGA show low rates of TLF, revascularization and ST. This supports safety and efficacy of the OMEGA BMS for the treatment of coronary artery disease.

OriģinālvalodaAngļu
Lapas (no-līdz)65-69
Lapu skaits5
ŽurnālsCardiovascular Revascularization Medicine
Sējums16
Izdevuma numurs2
DOIs
Publikācijas statussPublicēts - 2015
Ārēji publicēts

ANO IAM

Šis izpildes rezultāts palīdz sasniegt šādus ANO ilgtspējīgas attīstības mērķus (IAM)

  1. 3. IAM — Laba Veselība un Labbūtība
    3. IAM — Laba Veselība un Labbūtība

Nospiedums

Uzziniet vairāk par pētniecības tēmām “Primary endpoint results of the OMEGA Study: One-year clinical outcomes after implantation of a novel platinum chromium bare metal stent”. Kopā tie veido unikālu nospiedumu.

Citēt šo