Stent underexpansion is associated with high wall shear stress: a biomechanical analysis of the shear stent study

  • Sonali Kumar
  • , David Molony
  • , Sameer Khawaja
  • , Kaylyn Crawford
  • , Elizabeth W. Thompson
  • , Olivia Hung
  • , Imran Shah
  • , Jessica Navas-Simbana
  • , Arlen Ho
  • , Arnav Kumar
  • , Yi An Ko
  • , Hossein Hosseini
  • , Adrien Lefieux
  • , Joo Myung Lee
  • , Joo Yong Hahn
  • , Shao Liang Chen
  • , Hiromasa Otake
  • , Takashi Akasaka
  • , Eun Seok Shin
  • , Bon Kwon Koo
  • Goran Stankovic, Dejan Milasinovic, Chang Wook Nam, Ki Bum Won, Javier Escaned, Andrejs Ērglis, Yoshinobu Murasato, Alessandro Veneziani, Habib Samady
  • Institute of Cardiology and Regenerative Medicine
  • Emory University
  • Nanjing Medical University
  • Clinical Center of Serbia
  • Paula Stradina Clinical University Hospital

Zinātniskās darbības rezultāts: Devums žurnālamZinātniskais raksts (žurnālā)koleģiāli recenzēts

5 Atsauces (Scopus)

Kopsavilkums

Coronary stent underexpansion is associated with restenosis and stent thrombosis. In clinical studies of atherosclerosis, high wall shear stress (WSS) has been associated with activation of prothrombotic pathways, upregulation of matrix metalloproteinases, and future myocardial infarction. We hypothesized that stent underexpansion is predictive of high WSS. WSS distribution was investigated in patients enrolled in the prospective randomized controlled study of angulated coronary arteries randomized to undergo percutaneous coronary intervention with R-ZES or X-EES. WSS was calculated from 3D reconstructions of arteries from intravascular ultrasound (IVUS) and angiography using computational fluid dynamics. A logistic regression model investigated the relationship between WSS and underexpansion and the relationship between underexpansion and stent platform. Mean age was 63±11, 78% were male, 35% had diabetes, mean pre-stent angulation was 36.7°±14.7°. Underexpansion was assessed in 83 patients (6,181 IVUS frames). Frames with stent underexpansion were significantly more likely to exhibit high WSS (> 2.5 Pa) compared to those without underexpansion with an OR of 2.197 (95% CI = [1.233–3.913], p = 0.008). There was no significant association between underexpansion and low WSS (< 1.0 Pa) and no significant differences in underexpansion between R-ZES and X-EES. In the Shear Stent randomized controlled study, underexpanded IVUS frames were more than twice as likely to be associated with high WSS than frames without underexpansion.

OriģinālvalodaAngļu
Lapas (no-līdz)1375-1382
Lapu skaits8
ŽurnālsInternational Journal of Cardiovascular Imaging
Sējums39
Izdevuma numurs7
DOIs
Publikācijas statussPublicēts - jūl. 2023

ANO IAM

Šis izpildes rezultāts palīdz sasniegt šādus ANO ilgtspējīgas attīstības mērķus (IAM)

  1. 3. IAM — Laba Veselība un Labbūtība
    3. IAM — Laba Veselība un Labbūtība

OECD Zinātnes nozare

  • 3. Medicīnas un veselības zinātnes

Nospiedums

Uzziniet vairāk par pētniecības tēmām “Stent underexpansion is associated with high wall shear stress: a biomechanical analysis of the shear stent study”. Kopā tie veido unikālu nospiedumu.

Citēt šo