Kopsavilkums
In this work, we present a novel, compact, power efficient and variable magnetic field source design for magnetic field imaging microscopy. The device is based on a pair of diametrically magnetized permanent magnet cylinders with electro-mechanical rotation control and ferrite flux homogenizers. A Hall probe and NV centers in diamond are used to demonstrate a proof of concept of a proposed magnetic field setup and to characterize the homogeneity of the produced magnetic field on a micrometer scale. Numerical simulation results are compared with experimental results showing good agreement of the distribution of the magnetic field in the setup. As a result, a magnetic field source with a tunable field amplitude in the range from 1 mT to 222 mT is demonstrated, achieving a magnetic field homogeneity of 2 ppm/μm or 0.5 μT/μm at 222 mT in a 25 × 25 μm field of view.
| Oriģinālvaloda | Angļu |
|---|---|
| Raksta numurs | 113624 |
| Žurnāls | Ultramicroscopy |
| Sējums | 242 |
| DOIs | |
| Publikācijas statuss | Publicēts - dec. 2022 |
OECD Zinātnes nozare
- 1.3 Fizika un astronomija
Nospiedums
Uzziniet vairāk par pētniecības tēmām “Tunable magnetic field source for magnetic field imaging microscopy”. Kopā tie veido unikālu nospiedumu.Citēt šo
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver