Validation of a method for measuring the retinal thickness with Shack-Hartmann aberrometry in an artificial eye

  • University of Latvia

Zinātniskās darbības rezultāts: Devums žurnālam › Zinātniskais raksts (žurnālā) › koleģiāli recenzēts

Kopsavilkums

In Shack-Hartmann aberrometry, it is assumed that a wave front emerges from a single point focused on a retina. However, the retina is a multi-layered structure and reflections may occur from several layers. This may result in several overlapping spot patterns on the CCD due to different vergences of the outgoing wave fronts. The amount by which these spot patterns are displaced may contain information about the retinal thickness. In this study, we perform simulations of formation of double spots in a living eye and also apply this method to measure the thickness of an artificial retina with a simple structure. We also compare the results obtained with artificial eye and compare them to the simulated data. We evaluate the recommended range of the lenslet parameters for analyzing the retinal thickness. We conclude that this method could be used in a living eye for estimating the total retinal thickness and to confirm retinal pathologies associated with significant increase in the retinal thickness like glaucoma, macular edema, etc.

OriģinālvalodaAngļu
Lapas (no-līdz)662-675
Lapu skaits14
ŽurnālsJournal of Modern Optics
Sējums62
Izdevuma numurs8
DOIs
Publikācijas statussPublicēts - 4 maijs 2015

Nospiedums

Uzziniet vairāk par pētniecības tēmām “Validation of a method for measuring the retinal thickness with Shack-Hartmann aberrometry in an artificial eye”. Kopā tie veido unikālu nospiedumu.

Citēt šo